NANOCOLOR UV/VIS II, 190−1100 nm Spectrophotometer

NANOCOLOR UV/VIS II, 190−1100 nm Spectrophotometer

The NANOCOLOR VIS II and NANOCOLOR UV/VIS II are high-precision measurement instruments applicable in all areas of water and wastewater analysis. MACHEREY‑NAGEL revolutionizes the daily laboratory work with these two new spectrophotometers, combining premium high-tech instruments with outstanding usability. With their intuitive, icon-based menu guidance, these innovative photometers can be used like a smartphone or tablet. The clearly arranged, high-resolution touch screen display makes your daily measurement routine a real pleasure.

STANDARD FEATURES

Brand       NANOCOLOR

Type          Spectrophotometer with reference detector technology (RDT)

Light source                         Halogen lamp (visible range), Deuterium lamp (UV range)

Optical system                     Monochromator

Wavelength                           190–1100 nm

Wavelength accuracy            ± 1 nm

Wavelength resolution           0.1 nm

Wavelength reproducibility      <0.5 nm

Wavelength calibration              Automatic

Wavelength selection                  Automatic, barcode, manual

Scan speed                                    1 complete scan in less than 1 min

Spectral bandwidth                    < 2 nm

Photometric range                      ± 3.0 A in wavelength range 200–900 nm

Photometric accuracy                 0.005 A at 0.0–0.5 A; 1 % at 0.5–2.0 A

Photometric linearity                  < 0.5 % at ≤ 2 A; ≤ 1 % at > 2 A

NANOCOLOR

REF 919600.1